Brand: KN

Product origin: Dalian,China

Delivery time: deliver as soon as received payment

Supply capacity: 30 sets one month

GPC gel purification system has many advantages such as high purification rate, simple operation, convenient maintenance, safety and reliability. It is mainly used in the pre-treatment and purification process of samples for the detection of pesticide and veterinary drug residues, polycyclic aromatic hydrocarbons, and polychlorinated biphenyls. It is widely used in the fields of agricultural products, food, environment and life sciences.

Category:

Description

KN-300 Gel Permeation Chromatography (GPC)

Overview

As we have seen, GPC/SEC is a type of LC and so solid stationary and liquid mobile phases are again used. However, the separation mechanism here relies solely on the size of the polymer molecules in solution, rather than any chemical interactions between particles and the stationary phase.

Features:

  1. High-precision high-pressure constant-flow pump, stable infusion, good repeatability
  2. Real-time pressure display, high and low pressure alarms and other functions to ensure the safety of the instrument
  3. Variable wavelength ultraviolet-visible detector, wide wavelength range, high sensitivity
  4. Touch screen automatic fraction collector, easy to operate and intuitive
  5. Stainless steel and quartz glass gel purification columns are optional, with good consistency, shortening operation time and improving work efficiency
  6. Floating self-calibration sealing ring technology and patented technology of pump head self-cleaning function, which can extend the service life of the sealing ring and ensure the durability and high efficiency of the instrument
  7. Convenient and easy-to-use flow and pressure correction function to ensure the high precision and stability of the instrument
  8. Adopt two kinds of gradient running program design: linear gradient and step gradient are optional
  9. Programmable solvent type and compression factor, which can automatically compensate for the complete pre-cleaning function, which is convenient for quick solvent replacement
  10. Adopts high-precision photodiode design, linear range expansion
  11. Convenient and unique light intensity gain coefficient self-correction function
  12. Unique holmium glass cell design, full wavelength scanning, convenient and accurate wavelength correction
  13. Universal sample cell design, easy to change between micro and constant cell

Configuration:

No. Name Specification Unit Qty.
 1  GPC High pressure infusion pump          1set

Variable wavelength ultraviolet detector  1set

Large volume manual injection valve    1pcs

Gel purification column (filled with imported gel filler to meet the pre-treatment of food or environmental samples)                1pcs

Chromatography workstation         1set

Manual fraction collector      1pcs

set 1

Technical parameters

High pressure constant flow pump
Measuring range 0.001-10ml/min (Step size setting : 0.001mL/min)
Max working pressure 42.0MPa Upper and lower limits can be set, automatic alarm
Flow accuracy ±1.0% or 2uL/min(0.001-10mL/min)
Flow precision 0.25%
Gradient Double
Gradient method Gradient elution internal software can be self-controlled, and can be controlled by PC and pump
Rated voltage AC 110V/220V,50Hz/60Hz
Dimension 400mm×300mm×180mm
Detector
Lamp source Deuterium Lamp / Tungsten Lamp
Wavelength range 190-700nm, increment : 1nm
Bandwidth <6m
Wavelength accuracy ±1nm
Wavelength repeatability 0.5nm
0.0005AU/mV-4AU/mV
Noise ±5×10-5AU
Drift 2×10-4AU(P-P)
Sample cell Analysis, Semi-prepared is optional

Application

Food:

America:AOAC 984.21 、FDA2905A

Europe:EN12393、EN1528、S-19、S35L00.00-34

China:GBT 5009.19-2008,GBT 5009.146-2008,GBT 5009.162-2008,GBT 5009.207-2008,GBT 5009.218-2008,GBT 19650-2006,SN/T2430-2010  ,SN/T 2457-2010  ,SN/T 2459-2010  ,SN/T 2431-2010  ,GB/T20772-2006

Medicine:

2015 Pharmacopoeia 2341: Determination of residues of some pesticides in medicinal materials, decoction pieces and preparations, determination methods of organochlorine and organophosphorus pesticide residues